Skip to content

ATTACH_ANALOG_INPUT_STRAIN_GAGE

Download Flojoy Studio to try this app
Attach channel(s) to a task measure strain. **Compatibility:** Compatible with National Instruments devices that utilize NI-DAQmx. Tested on a simulated NI-9236 module. This block is designed for use with Windows and Linux systems due to NI driver availability. Ensure you have installed the NI-DAQmx runtime from https://www.ni.com/en/support/downloads/drivers/download.ni-daq-mx.html. Params: task_name : str The name of the task to create. cDAQ_start_channel : NIDAQmxDevice The device and channel to read from. cDAQ_end_channel : NIDAQmxDevice To read from only one channel, set this to the same as `cDAQ_start_channel`. To read from multiple channels, set this to the last channel you want to read from. min_val : float Specifies in **units** the minimum value you expect to measure (default is -0.001). max_val : float Specifies in **units** the maximum value you expect to measure (default is 0.001). units : Literal The units to use to return strain measurements (default is "STRAIN"). strain_config : Literal Specifies information about the bridge configuration and measurement (default is "Full Bridge 1"). voltage_excitation_source : Literal Specifies the source of excitation (default is "Internal"). voltage_excitation_value : float Specifies in volts the amount of excitation to supply to the bridge (default is 2.5). gage_factor : float Contains information about the strain gage and measurement (default is 2.0). initial_bridge_voltage : float Specifies the initial voltage across the bridge before any strain is applied (default is 0.0). nominal_gage_resistance : float Contains information about the strain gage and measurement (default is 350.0). poisson_ratio : float Contains information about the strain gage and measurement (default is 0.3). lead_wire_resistance : float Specifies the resistance of the lead wires connected to the strain gage (default is 0.0). Returns: out : Optional[DataContainer] This block does not return any meaningful data; it is designed for creating a task to measure strain.
Python Code
from flojoy import flojoy, DataContainer, NIDAQmxDevice, DeviceConnectionManager
from typing import Optional, Literal
import nidaqmx


@flojoy(deps={"nidaqmx": "0.9.0"})
def ATTACH_ANALOG_INPUT_STRAIN_GAGE(
    task_name: str,
    cDAQ_start_channel: NIDAQmxDevice,
    cDAQ_end_channel: NIDAQmxDevice,
    min_val: float = -0.001,
    max_val: float = 0.001,
    units: Literal["STRAIN"] = "STRAIN",
    strain_config: Literal[
        "Full brige 1",
        "Full brige 2",
        "Full brige 3",
        "Half brige 1",
        "Half brige 2",
        "Quarter brige 1",
        "Quarter brige 2",
    ] = "Full brige 1",
    voltage_excitation_source: Literal["None", "External", "Internal"] = "Internal",
    voltage_excitation_value: float = 2.5,
    default: Optional[DataContainer] = None,
    gage_factor: float = 2.0,
    initial_bridge_voltage: float = 0.0,
    nominal_gage_resistance: float = 350.0,
    poisson_ratio: float = 0.3,
    lead_wire_resistance: float = 0.0,
) -> Optional[DataContainer]:
    """Attach channel(s) to a task measure strain.

    **Compatibility:**
    Compatible with National Instruments devices that utilize NI-DAQmx. Tested on a simulated NI-9236 module.

    This block is designed for use with Windows and Linux systems due to NI driver availability. Ensure you have installed the NI-DAQmx runtime from https://www.ni.com/en/support/downloads/drivers/download.ni-daq-mx.html.

    Parameters
    ----------
    task_name : str
        The name of the task to create.
    cDAQ_start_channel : NIDAQmxDevice
        The device and channel to read from.
    cDAQ_end_channel : NIDAQmxDevice
        To read from only one channel, set this to the same as `cDAQ_start_channel`. To read from multiple channels, set this to the last channel you want to read from.
    min_val : float, optional
        Specifies in **units** the minimum value you expect to measure (default is -0.001).
    max_val : float, optional
        Specifies in **units** the maximum value you expect to measure (default is 0.001).
    units : Literal, optional
        The units to use to return strain measurements (default is "STRAIN").
    strain_config : Literal, optional
        Specifies information about the bridge configuration and measurement (default is "Full Bridge 1").
    voltage_excitation_source : Literal, optional
        Specifies the source of excitation (default is "Internal").
    voltage_excitation_value : float, optional
        Specifies in volts the amount of excitation to supply to the bridge (default is 2.5).
    gage_factor : float, optional
        Contains information about the strain gage and measurement (default is 2.0).
    initial_bridge_voltage : float, optional
        Specifies the initial voltage across the bridge before any strain is applied (default is 0.0).
    nominal_gage_resistance : float, optional
        Contains information about the strain gage and measurement (default is 350.0).
    poisson_ratio : float, optional
        Contains information about the strain gage and measurement (default is 0.3).
    lead_wire_resistance : float, optional
        Specifies the resistance of the lead wires connected to the strain gage (default is 0.0).

    Returns
    -------
    Optional[DataContainer]
        This block does not return any meaningful data; it is designed for creating a task to measure strain.
    """
    # Check if task already exists
    task = DeviceConnectionManager.get_connection(task_name).get_handle()

    # Attach the requested channel(s) to the task
    name, address = cDAQ_start_channel.get_id().split("/")
    if cDAQ_end_channel:
        _, address_end = cDAQ_end_channel.get_id().split("/")
        address = f"{address}:{address_end[2:]}"
    physical_channels = f"{name}/{address}"

    units = nidaqmx.constants.StrainUnits.STRAIN  # TODO: Support custom scale
    strain_config = {
        "Full brige 1": nidaqmx.constants.StrainGageBridgeType.FULL_BRIDGE_I,
        "Full brige 2": nidaqmx.constants.StrainGageBridgeType.FULL_BRIDGE_II,
        "Full brige 3": nidaqmx.constants.StrainGageBridgeType.FULL_BRIDGE_III,
        "Half brige 1": nidaqmx.constants.StrainGageBridgeType.HALF_BRIDGE_I,
        "Half brige 2": nidaqmx.constants.StrainGageBridgeType.HALF_BRIDGE_II,
        "Quarter brige 1": nidaqmx.constants.StrainGageBridgeType.QUARTER_BRIDGE_I,
        "Quarter brige 2": nidaqmx.constants.StrainGageBridgeType.QUARTER_BRIDGE_II,
    }[strain_config]

    voltage_excitation_source = {
        "None": nidaqmx.constants.ExcitationSource.NONE,
        "External": nidaqmx.constants.ExcitationSource.EXTERNAL,
        "Internal": nidaqmx.constants.ExcitationSource.INTERNAL,
    }[voltage_excitation_source]

    task.ai_channels.add_ai_strain_gage_chan(
        physical_channels,
        min_val=min_val,
        max_val=max_val,
        units=units,
        strain_config=strain_config,
        voltage_excit_source=voltage_excitation_source,
        voltage_excit_val=voltage_excitation_value,
        gage_factor=gage_factor,
        initial_bridge_voltage=initial_bridge_voltage,
        nominal_gage_resistance=nominal_gage_resistance,
        poisson_ratio=poisson_ratio,
        lead_wire_resistance=lead_wire_resistance,
    )

Find this Flojoy Block on GitHub

Example App

Having problems with this example app? Join our Discord community and we will help you out!
React Flow mini map

This app shows how to use a strain gage input block for with a National Instruments Compact DAQ module.

Blocks used:

  • CREATE_TASK_ANALOG_INPUT_GAUGE
  • CONFIG_TASK_SAMPLE_CLOCK_TIMINGS
  • READ_TASK
  • LINE