EXTRACT_TRACE_2000

Download Flojoy Studio to try this app
Extract scope traces from a P2000 PicoScope. Note that this block requires the CONNECTION_2000 block. Also note that the P2000 block require a device specific driver/SDK downloaded from: https://www.picotech.com/downloads. Params: channel : select Select the channel to extract a trace from. coupling_type : select Select the coupling type (0:AC, 1:DC) trig_threshold : float The threshold to trigger the oscilloscope timing, in mV. trig_direction : select The direction (slope) to search for a trigger (1:Rising, 2:Falling). x_samples : int The number of samples in time (i.e. # of x axis pixels). Returns: out : OrderedPair The trace of the oscilloscope is returned.
Python Code
from flojoy import flojoy, DataContainer, OrderedPair
from typing import Optional, Literal
import ctypes
from numpy import linspace
from picosdk.ps2000 import ps2000 as ps
from picosdk.functions import adc2mV, assert_pico2000_ok, mV2adc


@flojoy(deps={"picosdk": "1.1"})
def EXTRACT_TRACE_2000(
    channel: Literal["A", "B"] = "A",
    coupling_type: Literal["0", "1", "2"] = "1",
    trig_threshold: float = 300,
    trig_direction: Literal["0", "1", "2"] = "1",
    x_samples: int = 1000,
    default: Optional[DataContainer] = None,
) -> OrderedPair:
    """Extract scope traces from a P2000 PicoScope.

    Note that this block requires the CONNECTION_2000 block.

    Also note that the P2000 block require a device specific driver/SDK downloaded from:
    https://www.picotech.com/downloads.

    Parameters
    ----------
    channel: select
        Select the channel to extract a trace from.
    coupling_type: select
        Select the coupling type (0:AC, 1:DC)
    trig_threshold: float
        The threshold to trigger the oscilloscope timing, in mV.
    trig_direction: select
        The direction (slope) to search for a trigger (1:Rising, 2:Falling).
    x_samples: int
        The number of samples in time (i.e. # of x axis pixels).

    Returns
    -------
    OrderedPair
        The trace of the oscilloscope is returned.
    """

    status = {}
    device_num = ctypes.c_int16(1)
    # find maximum ADC count value
    maxADC = ctypes.c_int16(32767)

    # Set up channel
    if channel == "A":
        chan_num = 0  # 0=A, 1=B
    else:
        chan_num = 1
    enabled = 1
    coupling_type = int(coupling_type)
    ch_range = 7
    status["setCh"] = ps.ps2000_set_channel(
        device_num, chan_num, enabled, coupling_type, ch_range
    )
    assert_pico2000_ok(status["setCh"])

    # Set up single trigger
    source = chan_num
    # convert mV counts data to ADC
    threshold = mV2adc(trig_threshold, ch_range, maxADC)
    direction = int(trig_direction)  # 0=RISING 1=falling
    delay = 0  # ms
    trigger = 1000  # ms
    status["trigger"] = ps.ps2000_set_trigger(
        device_num, source, int(threshold / 16), direction, delay, trigger
    )
    assert_pico2000_ok(status["trigger"])

    # Get timebase information
    timebase = 8
    timeInterval = ctypes.c_int32()
    timeUnits = ctypes.c_int32()
    oversample = ctypes.c_int16(1)
    maxSamplesReturn = ctypes.c_int32()
    status["getTimebase"] = ps.ps2000_get_timebase(
        device_num,
        timebase,
        x_samples,
        ctypes.byref(timeInterval),
        ctypes.byref(timeUnits),
        oversample,
        ctypes.byref(maxSamplesReturn),
    )
    assert_pico2000_ok(status["getTimebase"])

    # Run block capture
    timeIndisposedms = ctypes.c_int32()
    status["runBlock"] = ps.ps2000_run_block(
        device_num, x_samples, timebase, oversample, ctypes.byref(timeIndisposedms)
    )
    assert_pico2000_ok(status["runBlock"])

    # Check for data collection to finish using ps5000aIsReady
    ready = ctypes.c_int16(0)
    check = ctypes.c_int16(0)
    while ready.value == check.value:
        status["isReady"] = ps.ps2000_ready(device_num)
        ready = ctypes.c_int16(status["isReady"])

    # Create buffers ready for data
    bufferA = (ctypes.c_int16 * x_samples)()
    bufferB = (ctypes.c_int16 * x_samples)()

    # Get data from scope
    cmaxSamples = ctypes.c_int32(x_samples)
    status["getValues"] = ps.ps2000_get_values(
        device_num,
        ctypes.byref(bufferA),
        ctypes.byref(bufferB),
        None,
        None,
        ctypes.byref(oversample),
        cmaxSamples,
    )
    assert_pico2000_ok(status["getValues"])

    # convert ADC counts data to mV
    if channel == "A":
        signal = adc2mV(bufferA, ch_range, maxADC)
    else:
        signal = adc2mV(bufferB, ch_range, maxADC)

    # Create time data
    time = linspace(0, (cmaxSamples.value - 1) * timeInterval.value, cmaxSamples.value)

    # # Stop the scope
    # status["stop"] = ps.ps2000_stop(device_num)
    # assert_pico2000_ok(status["stop"])

    return OrderedPair(x=time, y=signal)