PROLOGIX_EOI
Download Flojoy Studio to try this app
Sets the EOI and EOS settings for the Prologix GPIB-USB adapter. Only used for the GPIB port (not the serial port). These settings depend on
the instrument you are using.
EOI - end of interrupt (use terminator or not)
EOS - end of string (termination character)
Requires an OPEN_SERIAL block. Params: connection : Serial The open serial connection with the instrument. eoi : bool Use EOI (1) or not (0). eos : select Which terminator to use. Returns: out : String Response from the Prologix USB-to-GPIB controller.
Python Code
import serial
from flojoy import flojoy, SerialConnection, DataContainer, String
from typing import cast, Optional, Literal
@flojoy(inject_connection=True)
def PROLOGIX_EOI(
connection: SerialConnection,
default: Optional[DataContainer] = None,
eoi: bool = True,
eos: Literal["CR+LF", "CR", "LF", "None"] = "None",
) -> String:
"""Sets the EOI and EOS settings for the Prologix GPIB-USB adapter.
Only used for the GPIB port (not the serial port). These settings depend on
the instrument you are using.
EOI - end of interrupt (use terminator or not)
EOS - end of string (termination character)
Requires an OPEN_SERIAL block.
Parameters
----------
connection: Serial
The open serial connection with the instrument.
eoi: bool
Use EOI (1) or not (0).
eos: select
Which terminator to use.
Returns
-------
String
Response from the Prologix USB-to-GPIB controller.
"""
# Start serial communication with the instrument
ser = cast(serial.Serial, connection.get_handle())
if ser is None:
raise ValueError("Serial communication is not open")
if eoi:
cmd = 1
else:
cmd = 0
match eos:
case "CR+LF":
eos = 0
case "CR":
eos = 1
case "LF":
eos = 2
case "None":
eos = 3
ser.write(str(cmd).encode())
cmd = f"++eos {eos}\n"
ser.write(cmd.encode())
s = ser.read(256)
return String(s)
Example App
Having problems with this example app? Join our Discord community and we will help you out!
This app prepares the Prologix GPIB-USB adapter for use with a VISA instrument. The user can then send a SCPI command to a VISA instrument using the GPIB-USB adapter.
Parameters:
- Connection on all of the Blocks should be the COM port of the adapter.
PROLOGIX AUTO
: auto = offPROLOGIX MODE
: mode = CONTROLLERPROLOGIX EOI
: EOI = true, EOS = None (instrument dependent)- GPIB address should match the instrument’s (see the instrument’s manual on how to find it)
OPEN SERIAL
: writing “*IDN?” with bytes encoding and a LF terminator
This app has two rows that are run seperately:
- The top, setup, row with
AUTO
,MODE
andEOI
blocks. - The bottom row with
WRITE
andREAD
.
The two constants on the bottom left control which row runs. If they are equal, the setup row runs. Otherwise, the read/write row runs. The setup should be run before connecting the adapter to the instrument.
Note that the PROLOGIX READ
should be used to recieve a response from the instrument. The AUTO mode can be turned on which recieves a response upon every command (PROLOGIX AUTO
block controls this setting). However, this causes errors in some VISA instruments.