PROLOGIX_READ
Download Flojoy Studio to try this app
Returns the response from the GPIB instrument. For example if you query a VISA instrument with "*IDN?" using a
SERIAL_WRITE block, this block must be used to return the response.
Not required if the PROLOGIX_AUTO setting is "on" (not recommended).
Requires an OPEN_SERIAL block. Params: connection : Serial The open serial connection with the instrument. Returns: out : String Response from the Prologix USB-to-GPIB controller.
Python Code
import serial
from flojoy import flojoy, SerialConnection, DataContainer, String
from typing import cast, Optional
@flojoy(inject_connection=True)
def PROLOGIX_READ(
connection: SerialConnection,
default: Optional[DataContainer] = None,
) -> String:
"""Returns the response from the GPIB instrument.
For example if you query a VISA instrument with "*IDN?" using a
SERIAL_WRITE block, this block must be used to return the response.
Not required if the PROLOGIX_AUTO setting is "on" (not recommended).
Requires an OPEN_SERIAL block.
Parameters
----------
connection: Serial
The open serial connection with the instrument.
Returns
-------
String
Response from the Prologix USB-to-GPIB controller.
"""
# Start serial communication with the instrument
ser = cast(serial.Serial, connection.get_handle())
if ser is None:
raise ValueError("Serial communication is not open")
cmd = "++read eoi\n"
ser.write(cmd.encode())
s = ser.read(256)
if isinstance(s, bytes):
s = s.decode("utf-8")
return String(s)
Example App
Having problems with this example app? Join our Discord community and we will help you out!
This app prepares the Prologix GPIB-USB adapter for use with a VISA instrument. The user can then send a SCPI command to a VISA instrument using the GPIB-USB adapter.
Parameters:
- Connection on all of the Blocks should be the COM port of the adapter.
PROLOGIX AUTO
: auto = offPROLOGIX MODE
: mode = CONTROLLERPROLOGIX EOI
: EOI = true, EOS = None (instrument dependent)- GPIB address should match the instrument’s (see the instrument’s manual on how to find it)
OPEN SERIAL
: writing “*IDN?” with bytes encoding and a LF terminator
This app has two rows that are run seperately:
- The top, setup, row with
AUTO
,MODE
andEOI
blocks. - The bottom row with
WRITE
andREAD
.
The two constants on the bottom left control which row runs. If they are equal, the setup row runs. Otherwise, the read/write row runs. The setup should be run before connecting the adapter to the instrument.
Note that the PROLOGIX READ
should be used to recieve a response from the instrument. The AUTO mode can be turned on which recieves a response upon every command (PROLOGIX AUTO
block controls this setting). However, this causes errors in some VISA instruments.